Flagship model of thickness tester with built-in printer
Must be used with an optional probe (EP-100 electromagnetic probe or HP-100 eddy current probe)
Measurement range of electromagnetic type: 0 - 2500 µm; eddy current type: 0 – 1,200 µm
Records 24,000 data sets, statistical calculation function
Large display
On the large display, the calibration procedure is displayed interactively and
statistical results are displayed visibly and clearly.
High visibility is provided for dark places thanks to the built-in backlight.
Specifications
Measurement Method
Electromagnetic induction/Eddy-current
Applications
Non-magnetic coating on iron and steel (ferrous) substrates and Insulating coatings on non-ferrous substrates
Measurement Range
*Depends on the probe connected
Electromagnetic (EP-100): 0-2,500µm or 99.0mils
Eddy-current (HP-100): 0-1,200µm or 47.0mils
Measurement Precision
*Depends on the probe connected
Electromagnetic (EP-100): Under 15µm: ±0.3mm, 15µm or greater: ±2%, 1,000μm or greater: ±3%
Eddy-current (HP-100): Under 50μm:±1.0µm, 50µm or greater: ±2%
Resolution
0.1µm (less than 100µm), 1µm (100µm or greater)
Conformity Standard
Electromagnetic induction: JIS K5600-1-7, JIS H0401, JIS H8401, JIS H8501, / ISO 2808, ISO 2064, ISO 1460, ISO 2178, ISO 19840 / ASTM B 499, ASTM D 7091, ASTM E 376
Eddy-current: JIS K5600-1-7, JIS H8680-2, JIS H8501 / ISO 2808, ISO 2360, ISO 2064, ISO 19840 /ASTM B 244, ASTM D 7091, ASTM E 376
Statistical Functions
Number of measurements, Average value, Standard deviation, Maximum value, Minimum value, Block and group calculation
Applicable Probe
One-point contact fixed pressure (EP-100, HP-100)
Display Format
Digital (LCD, smallest displayed unit 0.1mm)
Output
USB serial
Power Source
AC100V-220V or 1.5V "AA" size Alkaline batteries x 8
(x4 for main unit and x4 for printer)
Dimensions & Weight
126mm(W)×256mm(D)×93mm(H)750g
Accessories
Iron substrate (for Fe probe), aluminum substrate (for NFe probe), calibration foil x 6, battery 1.5V (AA alkali) ×8, AC adapter, power cable POC00, probe adapter, strap, printer paper ×2, surface protective sheet ×3, calibration foil case, carrying case, Easy Guide for Calibration, Operating Manual
Options
Calibration foils (thickness other than those supplied), Measuring Stand LW-990