Features
● Four ways to present waveform: positive half-wave, negative half- wave, full wave and radio frequency.
● Automatic gain adjustment, defect equivalent calculation and peak memory function
● Two high resolution scanning mode: A and B
● Display of echo envelope
● Two individual gates setting and alarming function.
● 32 detecting channels are available with separate detecting parameters and DAC (Distance Amplitude Correction) curves in every channel.
● Automatic formation of DAC curves, and 30 points’ data can be recorded infinitely, adjustable offset curves and gain correction functions are available.
●Three detecting modes(single-probe, dual crystal probe and transmission) with automatic calibration function
● Connected to PC via USB interface with advanced software for data analysis and management.
● Data and documents are managed with file allocation table (FAT)system, making the management of inspection data more convenient, reliant and faster
● Super large memory up to 32M, 1000 echo data can be stored.
● EL Highlight matrix display (no drift with angle, temperature or sunlight)
● Brand new digital signal circuit is designed for TUD310
● Digital signal processor (DSP) is used for signals analyzing, making circuit noise reduced properly and waveform more stable.
● EPSON ink-jet printers can be connected with TUD
● Real-time waveform display and review
Specification
Items
|
Description
|
Scanning range
|
2.5 mm ~9999 mm
|
Scanning resolution
|
0.1mm (2.5mm ~100mm)
1mm (100 mm ~5000mm)
|
Gain range
|
0dB ~110 dB
|
D-delay
|
-20µs~+3400µs
|
P-delay
|
0µs~99.99µs,resolution 0.01µs
|
Sound speed
|
1000 m/s~9999m/s
|
Bandwidth
|
0.2MHz~15MHz
(Low0.2~1 Mid.0.5~4 High3~15)
|
Vertical linearity accuracy
|
≤3%
|
Horizontal linearity accuracy
|
≤0.2%
|
Dynamic range
|
≥32dB
|
Rectification
|
Positive half wave, negative wave, full wave, and RF
|
Sensitivity leavings
|
≥60dB
|
Test mode
|
Pulse-echo, dual and through transmission
|
Pulser
|
Spike excitation pulser
|
Damping
|
50ohms, 150ohms and 400ohms
|
Reject
|
Linear, 0-80% of full screen, variable in steps of 1%
|
Unit
|
Metric/inch
|
Interface
|
RS232 / USB
|
Printer
|
EPSON ink-jet printers
|
AC requirements
|
85-264V AC/1.0A,47-63Hz
|
Temperature
|
-10℃~40℃
|
Humidity
|
20%~90%RH
|
Power supply
|
Li battery4×3.6V 4000mAh
|
Charging time
|
4~5hours
|
Dimension (mm)
|
243×173×70
|
Weight (g)
|
1470
|